Frequently Asked Questions
These are questions I've been asked about the PM3055.
- I've heard that there are
special probes for this scope that allow it to auto-adjust some settings
during operation.
My question is, do you use these probes and if yes, can you tell me what the
exact part number of these probes is?
- I'm just using regular off-the-shelf Welleman probes. The auto-adjust feature works no matter which probe is attached. I don't think there is anything gained by using special probes.
- What is the 'remote control' connector on the back for?
- It gives access to the internal I2C bus of the instrument. It also
allows you to externally control the auto set button by a single signal.
Here's the pin-out:
| 1 | SDA |
| 2 | +5V |
| 3 | SCL |
| 4 | GND |
| 5 | GND |
| 6 | +5V |
| 7 | FOOTN: Remote AUTO SET. Command line to initiate AUTO SET function. |
| 8 | GND |
| 9 | REMRQN: REMote ReQuest Not. Command line to initate transfer to/from oscilloscope. |
- Where can I get a manual?
- I do not know, but I guess that you could get it as a spare part from
your local Philips representative.
Expect to pay a lot for it.
I can scan individual pages for you if you ask me.
- How much is it worth?
- I wouldn't pay more than EUR 250 to 300 for one if the condition is good.
- Where can I find more information on the unit (e.g. repair information)?
- There is not much on the Internet (e.g. the
search for "PM3055" on Google!
doesn't give much), so I guess the only way you can find out more is by
buying maintenance and repair and users manuals from Philips. Or ask me, I
haven't actually tried repairing my scope, but I do have the maintenance
manaual.